Optical coherence tomography
This article has multiple issues. Please help improve it or discuss these issues on the talk page. (Learn how and when to remove these template messages)
|
Optical coherence tomography (OCT) is an imaging technique that uses
The first US patents on OCT by the MIT/Harvard group described both a time-domain OCT (TD-OCT) system[6][7] and a Fourier-domain OCT (FD-OCT) system of the swept-source variety.[6] These patents were licensed by Zeiss and formed the basis of the first generations of OCT products until 2006. Tanno et al. obtained a patent on optical heterodyne tomography (similar to TD-OCT) in Japan in the same year.[8]
In the decade preceding the invention of OCT, interferometry with short-coherence-length light had been investigated for a variety of applications.[9][10][11][12][13][14][15][16][17][18][19][20] The potential to use interferometry for imaging was proposed,[20] and measurement of retinal elevation profile and thickness had been demonstrated.[19]
The initial commercial clinical OCT systems were based on point-scanning TD-OCT technology, which primarily produced cross-sectional images due to the speed limitation (tens to thousands of axial scans per second). Fourier-domain OCT became available clinically 2006, enabling much greater image acquisition rate (tens of thousands to hundreds of thousands axial scans per second) without sacrificing signal strength. The higher speed allowed for three-dimensional imaging, which can be visualized in both en face and cross-sectional views. Novel contrasts such as angiography, elastography, and optoretinography also became possible by detecting signal change over time. Over the past three decades, the speed of commercial clinical OCT systems has increased more than 1000-fold, doubling every three years and rivaling Moore's law of computer chip performance. Development of parallel image acquisition approaches such as line-field and full-field technology may allow the performance improvement trend to continue.
OCT is most widely used in ophthalmology, in which it has transformed the diagnosis and monitoring of retinal diseases, optic nerve diseases, and corneal diseases. It has greatly improved the management of the top three causes of blindness – macular degeneration, diabetic retinopathy, and glaucoma – thereby preventing vision loss in many patients. By 2016 OCT was estimated to be used in more than 30 million imaging procedures per year worldwide.[21]
OCT angioscopy is used in the intravascular evaluation of coronary artery plaques and to guide stent placement. Beyond ophthalmology and cardiology, applications are also developing in other medical specialties such as dermatology, gastroenterology (endoscopy), neurology, oncology, and dentistry.[22][23]
Introduction
Interferometric reflectometry of biological tissue, especially of the human eye using short-coherence-length light (also referred to as partially-coherent, low-coherence, or broadband, broad-spectrum, or white light) was investigated in parallel by multiple groups worldwide since 1980s. In 1991, David Huang, then a student in James Fujimoto laboratory at Massachusetts Institute of Technology, working with Eric Swanson at the MIT Lincoln Laboratory and colleagues at the Harvard Medical School, successfully demonstrated imaging and called the new imaging modality "optical coherence tomography".[24] Since then, OCT with micrometer resolution and cross-sectional imaging capabilities has become a prominent biomedical imaging technique that has continually improved in technical performance and range of applications. The improvement in image acquisition rate is particularly spectacular, starting with the original 0.8 Hz axial scan repetition rate[1] to the current commercial clinical OCT systems operating at several hundred kHz and laboratory prototypes at multiple MHz. The range of applications has expanded from ophthalmology to cardiology and other medical specialties. For their roles in the invention of OCT, Fujimoto, Huang, and Swanson received the 2023 Lasker-DeBakey Clinical Medical Research Award and the National Medal of Technology and Innovation.[25] These developments have been reviewed in articles written for the general[21] scientific[26] and medical[27] readership.
It is particularly suited to ophthalmic applications and other tissue imaging requiring micrometer resolution and millimeter penetration depth.
OCT is based on low-coherence interferometry.
Light in an OCT system is broken into two arms – a sample arm (containing the item of interest) and a reference arm (usually a mirror). The combination of reflected light from the sample arm and reference light from the reference arm gives rise to an interference pattern, but only if light from both arms have traveled the "same" optical distance ("same" meaning a difference of less than a coherence length). By scanning the mirror in the reference arm, a reflectivity profile of the sample can be obtained (this is time domain OCT). Areas of the sample that reflect back a lot of light will create greater interference than areas that don't. Any light that is outside the short coherence length will not interfere.
Layperson's explanation
This section needs additional citations for verification. (March 2019) |
Optical coherence tomography (OCT) is a technique for obtaining sub-surface images of translucent or opaque materials at a resolution equivalent to a low-power microscope. It is effectively "optical ultrasound", imaging reflections from within tissue to provide cross-sectional images.[35]
OCT has attracted interest among the medical community because it provides tissue morphology imagery at much higher resolution (less than 10 μm axially and less than 20 μm laterally[36] ) than other imaging modalities such as MRI or ultrasound.
The key benefits of OCT are:
- Live sub-surface images at near-microscopic resolution
- Instant, direct imaging of tissue morphology
- No preparation of the sample or subject, no contact
- No ionizing radiation
OCT delivers high resolution because it is based on light, rather than sound or radio frequency. An optical beam is directed at the tissue, and a small portion of this light that reflects from sub-surface features is collected. Note that most light is not reflected but, rather, scatters off at large angles. In conventional imaging, this diffusely scattered light contributes background that obscures an image. However, in OCT, a technique called interferometry is used to record the optical path length of received photons allowing rejection of most photons that scatter multiple times before detection. Thus OCT can build up clear 3D images of thick samples by rejecting background signal while collecting light directly reflected from surfaces of interest.
Within the range of noninvasive three-dimensional imaging techniques that have been introduced to the medical research community, OCT as an echo technique is similar to
The technique is limited to imaging 1 to 2 mm below the surface in biological tissue, because at greater depths the proportion of light that escapes without scattering is too small to be detected. No special preparation of a biological specimen is required, and images can be obtained "non-contact" or through a transparent window or membrane. It is also important to note that the laser output from the instruments used is low – eye-safe near-infrared or visible-light[38] – and no damage to the sample is therefore likely.
Theory
The principle of OCT is white light, or low coherence, interferometry. The optical setup typically consists of an interferometer (Fig. 1, typically Michelson type) with a low coherence, broad bandwidth light source. Light is split into and recombined from reference and sample arms, respectively.
Time domain
In time domain OCT the path length of the reference arm is varied in time (the reference mirror is translated longitudinally). A property of low coherence interferometry is that interference, i.e. the series of dark and bright fringes, is only achieved when the path difference lies within the coherence length of the light source. This interference is called autocorrelation in a symmetric interferometer (both arms have the same reflectivity), or cross-correlation in the common case. The envelope of this modulation changes as path length difference is varied, where the peak of the envelope corresponds to path length matching.
The interference of two partially coherent light beams can be expressed in terms of the source intensity, , as
where represents the interferometer beam splitting ratio, and is called the complex degree of coherence, i.e. the interference envelope and carrier dependent on reference arm scan or time delay , and whose recovery is of interest in OCT. Due to the coherence gating effect of OCT the complex degree of coherence is represented as a Gaussian function expressed as[33]
where represents the spectral width of the source in the optical frequency domain, and is the centre optical frequency of the source. In equation (2), the Gaussian envelope is amplitude modulated by an optical carrier. The peak of this envelope represents the location of the microstructure of the sample under test, with an amplitude dependent on the reflectivity of the surface. The optical carrier is due to the Doppler effect resulting from scanning one arm of the interferometer, and the frequency of this modulation is controlled by the speed of scanning. Therefore, translating one arm of the interferometer has two functions; depth scanning and a Doppler-shifted optical carrier are accomplished by pathlength variation. In OCT, the Doppler-shifted optical carrier has a frequency expressed as
where is the central optical frequency of the source, is the scanning velocity of the pathlength variation, and is the speed of light.
The axial and lateral resolutions of OCT are decoupled from one another; the former being an equivalent to the coherence length of the light source and the latter being a function of the optics. The axial resolution of OCT is defined as
where and are respectively the central wavelength and the spectral width of the light source.[39]
Fourier domain
Fourier-domain (or Frequency-domain) OCT (FD-OCT) has speed and signal-to-noise ratio (SNR) advantages over time-domain OCT (TD-OCT) and has become the standard in the industry since 2006. The idea of using frequency modulation and coherent detection to obtain ranging information was already demonstrated in optical frequency domain reflectometry[9] and laser radar[14] in the 1980s, though the distance resolution and range were much longer than OCT. There are two types of FD-OCT – swept-source OCT (SS-OCT) and spectral-domain OCT (SD-OCT) – both of which acquire spectral interferograms which are then Fourier transformed to obtain an axial scan of reflectance amplitude versus depth. In SS-OCT, the spectral interferogram is acquired sequentially by tuning the wavelength of a laser light source. SD-OCT acquires spectral interferogram simultaneously in a spectrometer. An implementation of SS-OCT was described by the MIT group as early as 1994.[6][40] A group based in the University of Vienna described measurement of intraocular distance using both tunable laser and spectrometer-based interferometry as early as 1995.[41][42] SD-OCT imaging was first demonstrated both in vitro and in vivo by a collaboration between the Vienna group and a group based in the Nicholas Copernicus University in a series of articles between 2000 and 2002.[43][44][45] The SNR advantage of FD-OCT over TD-OCT was analyzed by multiple groups of researchers in 2003.[46][47][48]
Spectral-domain OCT
Spectral-domain OCT (spatially encoded frequency domain OCT) extracts spectral information by distributing different optical frequencies onto a detector stripe (line-array CCD or CMOS) via a dispersive element (see Fig. 4). Thereby the information of the full depth scan can be acquired within a single exposure. However, the large signal-to-noise advantage of FD-OCT is reduced due to the lower dynamic range of stripe detectors with respect to single photosensitive diodes, resulting in an SNR advantage of ~10 dB at much higher speeds. This is not much of a problem when working at 1300 nm, however, since dynamic range is not a serious problem at this wavelength range.[39]
The drawbacks of this technology are found in a strong fall-off of the SNR, which is proportional to the distance from the zero delay and a sinc-type reduction of the depth-dependent sensitivity because of limited detection linewidth. (One pixel detects a quasi-rectangular portion of an optical frequency range instead of a single frequency, the Fourier transform leads to the sinc(z) behavior). Additionally, the dispersive elements in the spectroscopic detector usually do not distribute the light equally spaced in frequency on the detector, but mostly have an inverse dependence. Therefore, the signal has to be resampled before processing, which cannot take care of the difference in local (pixelwise) bandwidth, which results in further reduction of the signal quality. However, the fall-off is not a serious problem with the development of new generation CCD or photodiode array with a larger number of pixels.
Synthetic array heterodyne detection offers another approach to this problem without the need for high dispersion.
Swept-source OCT
Swept-source OCT (Time-encoded frequency domain OCT) tries to combine some of the advantages of standard TD and spectral domain OCT. Here the spectral components are not encoded by spatial separation, but they are encoded in time. The spectrum is either filtered or generated in single successive frequency steps and reconstructed before Fourier transformation. By accommodation of a frequency scanning light source (i.e. frequency scanning laser) the optical setup (see Fig. 3) becomes simpler than spectral domain OCT, but the problem of scanning is essentially translated from the TD-OCT reference arm into the swept source OCT light source. Here the advantage lies in the proven high SNR detection technology, while swept laser sources achieve very small instantaneous bandwidths (linewidths) at very high frequencies (20–200 kHz). Drawbacks are the nonlinearities in the wavelength (especially at high scanning frequencies), the broadening of the linewidth at high frequencies and a high sensitivity to movements of the scanning geometry or the sample (below the range of nanometers within successive frequency steps).
Scanning schemes
Focusing the light beam to a point on the surface of the sample under test, and recombining the reflected light with the reference will yield an interferogram with sample information corresponding to a single A-scan (Z axis only). Scanning of the sample can be accomplished by either scanning the light on the sample, or by moving the sample under test. A linear scan will yield a two-dimensional data set corresponding to a cross-sectional image (X-Z axes scan), whereas an area scan achieves a three-dimensional data set corresponding to a volumetric image (X-Y-Z axes scan).
Single point
Systems based on single point, confocal, or flying-spot time domain OCT, must scan the sample in two lateral dimensions and reconstruct a three-dimensional image using depth information obtained by coherence-gating through an axially scanning reference arm (Fig. 2). Two-dimensional lateral scanning has been electromechanically implemented by moving the sample[49] using a translation stage, and using a novel micro-electro-mechanical system scanner.[50]
Line-field OCT
Line-field confocal optical coherence tomography (LC-OCT) is an imaging technique based on the principle of time-domain OCT with line illumination using a broadband laser and line detection using a line-scan camera.[51] LC-OCT produces B-scans in real-time from multiple A-scans acquired in parallel. En face as well as three-dimensional images can also be obtained by scanning the illumination line laterally.[52][53] The focus is continuously adjusted during the scan of the sample depth, using a high numerical aperture (NA) microscope objective to image with high lateral resolution. By using a supercontinuum laser as a light source, a quasi-isotropic spatial resolution of ~ 1 µm is achieved at a central wavelength of ~ 800 nm. On the other hand, line illumination and detection, combined with the use of a high NA microscope objective, produce a confocal gate that prevents most scattered light that does not contribute to the signal from being detected by the camera. This confocal gate, which is absent in the full-field OCT technique, gives LC-OCT an advantage in terms of detection sensitivity and penetration in highly scattering media such as skin tissues.[54] So far this technique has been used mainly for skin imaging in the fields of dermatology[55][56][57][58][59][60] and cosmetology.[61]
Full-field OCT
An imaging approach to temporal OCT was developed by Claude Boccara's team in 1998,[62] with an acquisition of the images without beam scanning. In this technique called full-field OCT (FF-OCT), unlike other OCT techniques that acquire cross-sections of the sample, the images are here "en-face" i.e. like images of classical microscopy: orthogonal to the light beam of illumination.[63]
More precisely, interferometric images are created by a Michelson interferometer where the path length difference is varied by a fast electric component (usually a piezo mirror in the reference arm). These images acquired by a CCD camera are combined in post-treatment (or online) by the phase shift interferometry method, where usually 2 or 4 images per modulation period are acquired, depending on the algorithm used.[64][65] More recently, approaches that allow rapid single-shot imaging were developed to simultaneously capture multiple phase-shifted images required for reconstruction, using single camera.[66] Single-shot time-domain OCM is limited only by the camera frame rate and available illumination.
The "en-face" tomographic images are thus produced by a wide-field illumination, ensured by the Linnik configuration of the Michelson interferometer where a microscope objective is used in both arms. Furthermore, while the temporal coherence of the source must remain low as in classical OCT (i.e. a broad spectrum), the spatial coherence must also be low to avoid parasitical interferences (i.e. a source with a large size).[67]
Selected applications
Optical coherence tomography is an established medical imaging technique and is used across several medical specialties including ophthalmology and cardiology, and is widely used in basic science research applications.
Ophthalmology
Ocular (or ophthalmic) OCT is used heavily by
Cardiology and neurovascular applications
In the settings of cardiology, OCT is used to image coronary arteries to visualize vessel wall lumen morphology and microstructure at a resolution ~10 times higher than other existing modalities such as intravascular ultrasounds, and x-ray angiography (intracoronary optical coherence tomography). For this type of application, 1 mm in diameter or smaller fiber-optics catheters are used to access artery lumen through semi-invasive interventions such as percutaneous coronary interventions.
The first demonstration of endoscopic OCT was reported in 1997, by researchers in James Fujimoto laboratory at Massachusetts Institute of Technology, including
Intravascular OCT has been further investigated for use in neurovascular applications including imaging for guiding endovascular treatment of ischemic stroke and brain aneurysms.[88] Clinical use of existing technologies has been limited to proximal intracranial anatomy of patient with limited tortuosity, however, it showed the potential of OCT for the imaging of neurovascular disease.[89] An intravascular OCT imaging catheter design tailored for use in tortuous neurovascular anatomy has been proposed in 2020.[90]
Further developments of intravascular OCT included the combination with other optical imaging modalities (multi-modality imaging). OCT has been combined with near-infrared fluorescence molecular imaging (NIRF) to enhance its capability to detect molecular/functional and tissue morphological information simultaneously.[91][92] In a similar way, combination with near-infrared spectroscopy (NIRS) has been implemented.[93]
Retinal imaging with PS-OCT demonstrated how the thickness and birefringence of blood vessel wall tissue of healthy subjects could be quantified, in vivo.[94] PS-OCT was subsequently applied to patients with diabetes and age-matched healthy subjects, and showed an almost 100% increase in vessel wall birefringence due to diabetes, without a significant change in vessel wall thickness.[95] In patients with hypertension however, the retinal vessel wall thickness increased by 60% while the vessel wall birefringence dropped by 20%, on average.[96] The large differences measured in healthy subjects and patients suggest that retinal measurements with PS-OCT could be used as a screening tool for hypertension and diabetes.[95][96][97]
Oncology
Endoscopic OCT has been applied to the detection and diagnosis of
Dermatology
The first use of OCT in dermatology dates back to 1997.[101] Since then, OCT has been applied to the diagnosis of various skin lesions including carcinomas.[102][103][104] However, the diagnosis of melanoma using conventional OCT is difficult, especially due to insufficient imaging resolution.[105] Emerging high-resolution OCT techniques such as LC-OCT have the potential to improve the clinical diagnostic process, allowing for the early detection of malignant skin tumors – including melanoma – and a reduction in the number of surgical excisions of benign lesions.[106] Other promising areas of application include the imaging of lesions where excisions are hazardous or impossible and the guidance of surgical interventions through identification of tumor margins.
Dentistry
Researchers in Tokyo medical and Dental University were able to detect enamel white spot lesions around and beneath the orthodontic brackets using swept source OCT.[107]
Research applications
Researchers have used OCT to produce detailed images of mice brains, through a "window" made of zirconia that has been modified to be transparent and implanted in the skull.
See also
- Angle-resolved low-coherence interferometry
- Ballistic photon
- Confocal microscopy
- Dual-axis optical coherence tomography
- Interferometry
- Intracoronary optical coherence tomography
- Leica Microsystems
- Medical imaging
- Novacam Technologies
- Optical heterodyne detection
- Optical projection tomography
- Spectroscopic optical coherence tomography
- Terahertz tomography
- Tomography
References
- ^ PMID 1957169.
- PMID 19738824.
- S2CID 121094027.
- PMID 19829430.
- PMID 8328536.
- ^ a b c US 5321501A, Swanson EA, Huang D, Fujimoto JG, Puliafito CA, "Method and apparatus for optical imaging with means for controlling the longitudinal range of the sample", issued 1994-06-14
- ^ US 5459570A, Swanson EA, Huang D, Fujimoto JG, Puliafito CA, "Method and apparatus for performing optical measurements", issued 1995-10-17
- ^ JA 2010042, Tanno N, Ichikawa T, Saeki A, "Lightwave Reflection Measurement", published 1990
- ^ ISSN 0003-6951.
- ISSN 0091-3286.
- S2CID 122883903.
- PMID 19738824.
- PMID 20454375.
- ^ S2CID 110631959.
- PMID 19742022.
- ISSN 1558-2213.
- S2CID 19888483.
- PMID 2001935.
- ^ ISBN 0-444-89860-3.
- ^ a b Shinji Chiba; Naohiro Tanno (1991). Backscattering Optical Heterodyne Tomography. 14th Laser Sensing Symposium (in Japanese).
- ^ PMID 27409459.
- PMID 26242713.
- PMID 27419359.
- PMID 1957169.
- PMID 37732757.
- PMID 37732756.
- S2CID 262085371.
- S2CID 20621284.
- PMID 11283681.
- PMID 15019397.
- PMID 11016028.
- ISBN 978-0-521-78449-8.
- ^ PMID 19742022.
- PMID 10933065.
- PMID 26618332.
- PMID 18073990.
- ^ "Optical Coherence Tomography provides better resolution than an MRI and Helps Diagnose Retina & Corneal Disease and Glaucoma, Part II". mastereyeassociates.com. mastereyeassociates. June 13, 2017. Retrieved June 13, 2017.
- PMID 29218923.
- ^ a b Garg A (2014). Anterior & Posterior Segment OCT: Current Technology & Future Applications, 1st edition.
- PMID 18183195.
- ISSN 0030-4018.
- PMID 18259516.
- PMID 18064195.
- S2CID 40844236.
- PMID 18026464.
- PMID 19461802.
- PMID 19466106.
- PMID 14587817.
- .
- .
- PMID 30650800.
- PMID 32206413.
- S2CID 234456595.
- PMID 17632613.
- PMID 30353716.
- S2CID 235232158.
- S2CID 230583854.
- S2CID 235411841.
- S2CID 236452537.
- S2CID 237241412.
- S2CID 208622348.
- PMID 18084473.
- PMID 17026940.
- S2CID 120323507.
- S2CID 123478275.
- PMID 33680554.
- ISBN 9781118574386.
- ^ "The ABCs of OCT". Review of Optometry.
- PMID 19465337.
- ^ "Outer Retinal Layers as Predictors of Vision Loss". Review of Ophthalmology.
- PMID 29037595.
- S2CID 39039199.
- PMID 22898648.
- PMID 25564068.
- PMID 21494659.
- S2CID 252564095.
- ^ Aik Kah T (2018). "CuRRL Syndrome: A Case Series" (PDF). Acta Scientific Ophthalmology. 1: 9–13.
- PMID 28760677.
- PMID 24614147.
- PMID 26599241.
- PMID 34393183.
- PMID 34393183.
- S2CID 43035300.
- PMID 19356512.
- ^ "LightLab launches FD-OCT in Europe" (Press release). Retrieved 9 September 2016.
- S2CID 258115402.
- ^ Swanson E (13 June 2016). "Optical Coherence Tomography: Beyond better clinical care: OCT's economic impact". BioOptics World. Retrieved 9 September 2016.
- PMID 29491139.
- S2CID 211159079.
- PMID 32737314.
- PMID 26971006.
- PMID 26685129.
- PMID 24514658.
- PMID 34457418.
- ^ PMID 37627295.
- ^ ISSN 0143-8166.
- ^ "Basic eye test could diagnose diabetes and hypertension". PerthNow. 2023-09-20. Retrieved 2024-01-24.
- ^ "Next-gen OCT for the esophagus". BioOptics World. 1 May 2013. Retrieved 9 September 2016.
- PMID 23314056.
- PMID 26977350.
- S2CID 20078741.
- S2CID 24965088.
- S2CID 26084419.
- PMID 25904111.
- PMID 28886803.
- S2CID 53023955. This article contains quotations from this source, which is available under the Creative Commons Attribution 3.0 Unported (CC BY 3.0)license.
- PMID 30158348.
- S2CID 14212180.
- ^ Mohan G (September 4, 2013). "A window to the brain? It's here, says UC Riverside team". Los Angeles Times.
- ^ US patent 7116429, Walecki WJ, Van P, "Determining thickness of slabs of materials", issued 2006-10-03.
- S2CID 120257179.
- S2CID 110785119.
- .
- S2CID 123249666.
- ^ Guss G, Bass I, Hackel R, Demos SG (November 6, 2007). High-resolution 3-D imaging of surface damage sites in fused silica with Optical Coherence Tomography (PDF) (Report). Lawrence Livermore National Laboratory. UCRL-PROC-236270. Archived from the original (PDF) on February 11, 2017. Retrieved December 14, 2010.
- ^ Walecki W, Wei F, Van P, Lai K, Lee T, Lau SH, Koo A (2004). Interferometric Metrology for Thin and Ultra-Thin Compound Semiconductor Structures Mounted on Insulating Carriers (PDF). CS Mantech Conference.
- ^ ISSN 2689-9620.
- S2CID 121209439.
- ^ See, for example: "ZebraOptical Optoprofiler: Interferometric Probe".
- ^ EP application 2799842, Markl, Daniel; Hannesschläger, Günther & Leitner, Michael et al., "A device and a method for monitoring a property of a coating of a solid dosage form during a coating process forming the coating of the solid dosage form", published 2014-11-05; GB application 2513581; US application 20140322429 A1.
- S2CID 120168355. Archived from the original(PDF) on 20 August 2011. Retrieved 25 May 2011.
- S2CID 120476700. Retrieved December 15, 2010.
- S2CID 15657288.
- .
- PMID 26968386.
- PMID 25266471.
- .
- PMID 33402664.