Crossed molecular beam
In
Technique
In a crossed molecular beam apparatus, two
History
The crossed molecular beam technique was developed by
Early crossed beam experiments investigated
Detecting scattered particles through a metal filament gave a good indication of angular distribution but has no sensitivity to kinetic energy. In order to gain insight into the kinetic energy distribution, early crossed molecular beam apparatuses used a pair of slotted disks placed between the collision center and the detector. By controlling the rotation speed of the disks, only particles with a certain known velocity could pass through and be detected.[2] With information about the velocity, angular distribution, and identity of the scattered species, useful information about the dynamics of the system can be derived.
Later improvements included the use of quadrupole mass filters to select only the products of interest,[5] as well as time-of-flight mass spectrometers to allow easy measurement of kinetic energy. These improvements also allowed the detection of a vast array of compounds, marking the advent of the "universal" crossed molecular beam apparatus.
The inclusion of supersonic nozzles to collimate the gases expanded the variety and scope of experiments, and the use of lasers to excite the beams (either before impact or at the point of reaction) further broadened the applicability of this technique.[2]