Radiation Exposure Monitoring
This article needs additional citations for verification. (June 2017) |
Radiation Exposure Monitoring (REM) is a framework developed by
Ready access to dose information aids medical staff, including
Collecting and using dose data
A challenge in automating the reporting of radiation exposure estimations has traditionally been a function of whether the record of dose provided by a manufacturer is persistent (i.e. stored electronically) or transient (i.e. displayed on a read-out). Many current radiology devices provide only transient records, either in the form of human-readable dose screens that require manual intervention (i.e. pencil and paper) to permanently capture the patient exposure, or else in the equally perishable data generated by a modality-performed procedure step (MPPS) created to help manage the scheduling system.[citation needed]
MPPS is insufficient, having a limited ability to encode complex data, and no options for long-term storage or queries. Newer scanners are able to create DICOM radiation dose structured reports (RDSRs) alongside the images themselves. REM addresses perishable dose data by creating a persistent record that can be sent to a central repository, and then queried and analyzed by health information systems for either a specific patient's history or for analysis of radiation exposure levels among patient groups, platforms, or clinical operations.[2] RDSRs, and the use of the IHE REM framework are part of the IEC 61910 standard.[3]
Standards and Integrating the Healthcare Enterprise (IHE)
The REM Profile enables imaging modalities to export radiation exposure estimation details in a standard format. Radiation reporting systems can either query for these "dose objects" periodically from an archive, or receive them directly from the modalities. The radiation reporting system is expected to perform relevant dose QA analysis and produce related reports. The analysis methods and report format are not considered topics for standardization and are not covered in the profile. The profile also describes how radiation reporting systems can submit dose estimation reports to centralized registries such as might be run by professional societies or national accreditation groups. In the United States, the American College of Radiology DIR[5] is one such registry. By profiling automated methods, the profile allows dose information to be collected and evaluated without imposing a significant administrative burden on staff otherwise occupied with caring for patients.
In addition to supporting profile quality assurance (QA) of the technical process at the local facility, (e.g. determining if the dose was appropriate for the procedure performed), the profile also supports population analysis performed by national registries. Compliant software products are capable of de-identifying and submitting dose reports to a national dose register securely, making it relatively simple for groups such as ACR to collect and process dose data from across the country once they have recruited participating sites.
Challenges
Fluoroscopy monitoring
Most
Limitations of dose monitoring
According to IHE, "It is important to understand the technical and practical limitations of dose monitoring and the reasons why the monitored values may not accurately provide the radiation dose administered to the patient":[1]
- The values provided by this tool are not "measurements" but only calculated estimates.
- For computed tomography, "CTDI" is a dose estimate to a standard plastic phantom. Plastic is not human tissue. Therefore, the dose should not be represented as the dose received by the patient.
- For planar or projection imaging, the recorded values may be exposure, skin dose or some other value that may not be patient's body or organ dose.
- It is inappropriate and inaccurate to add up dose estimates received by different parts of the body into a single cumulative value.
Despite such limitations, interest in monitoring radiation dose estimates is clearly expressed in such documents as the European directive Euratom 97/43[7] and the American College of Radiology Dose Whitepaper.[8]
References
- ^ a b "Radiation Exposure Monitoring". IHE Wiki. 7 March 2015. Retrieved 1 June 2017.
- S2CID 31939028.
- ^ "IEC 61910-1:2014". International Electrotechnical Commission. 24 September 2014. Retrieved 1 June 2017.
- ^ "Profiles". IHE. Retrieved 1 June 2017.
- ^ American College of Radiology (ACR) Dose Index Registry (DIR)
- ^ The Joint Commission FAQ Page
- ^ Eurotom 97/43 Archived 2012-10-30 at the Wayback Machine
- ^ ACR White Paper on Radiation Dose in Medicine
Further reading
- Thrall: Healthcare IT can help rads tackle radiation exposure
- Radiation reporting tools stalled in uncertain Euro markets
- California radiation fix-up bill adds dose of controversy
- White Paper - Radiation Exposure Monitoring
- Monitoring Radiation Dose: A New IHE Profile
- Google Group - IHE Radiology Technical Committee
- California's Dose Puzzle Is Radiology's Challenge, July 2012