Respirometry

Source: Wikipedia, the free encyclopedia.

Respirometry is a general term that encompasses a number of techniques for obtaining estimates of the rates of

microorganisms via an indirect measure of heat production (calorimetry
).

Whole-animal metabolic rates

The metabolism of an animal is estimated by determining rates of

thermic effect of food).[2] VO2max is typically determined during aerobic exercise at or near physiological limits.[3] In contrast, field metabolic rate (FMR) refers to the metabolic rate of an unrestrained, active animal in nature.[4] Whole-animal metabolic rates refer to these measures without correction for body mass. If SMR or BMR values are divided by the body mass value for the animal, then the rate is termed mass-specific. It is this mass-specific value that one typically hears in comparisons among species.[5]

Closed respirometry

Respirometry depends on a "what goes in must come out" principle.

hypoxic
.

Open respirometry

For an open-system, design constraints include washout characteristics of the animal chamber and sensitivity of the gas analyzers.

mammals
), chamber sizes and or flow rates would be selected to accommodate their higher metabolic rates.

Calculations

Calculating rates of VO2 and/or VCO2 requires knowledge of the flow rates into and out of the chamber, plus fractional concentrations of the gas mixtures into and out of the animal chamber. In general, metabolic rates are calculated from steady-state conditions (i.e., animal's metabolic rate is assumed to be constant[9][10]). To know the rates of oxygen consumed, one needs to know the location of the flow meter relative to the animal chamber (if positioned before the chamber, the flow meter is "upstream," if positioned after the chamber, the flow meter is "downstream"), and whether or not reactive gases are present (e.g., CO2, water, methane, see inert gas).

For an open system with upstream flow meter, water (e.g., anhydrous calcium sulfate) and CO2 removed prior to the oxygen analyzer, a suitable equation is

For an open system with downstream flow meter, water and CO2 removed prior to the oxygen analyzer, a suitable equation is

where

  • FR is the volumetric flow rate adjusted to STP (see
    Standard conditions for temperature and pressure
    )
  • FinO2 is the fractional amount of oxygen present in the incurrent air stream (the baseline or reference), and
  • FexO2 is the fractional amount of oxygen present in the excurrent air stream (what the animal has consumed relative to baseline per unit time).

For example, values for BMR of a 20 g mouse (

Mus musculus) might be FR = 200 mL/min, and readings of fractional concentration of O2 from an oxygen analyzer are FinO2 = 0.2095, FexO2 = 0.2072. The calculated rate of oxygen consumption is 0.58 mL/min or 35 mL/hour. Assuming an enthalpy of combustion for O2 of 20.1 joules
per milliliter, we would then calculate the heat production (and therefore metabolism) for the mouse as 703.5 J/h.

Respirometry equipment

For open flow system, the list of equipment and parts is long compared to the components of a closed system, but the chief advantage of the open system is that it permits continuous recording of metabolic rate. The risk of hypoxia is also much less in an open system.

Pumps for air flow

  • Vacuum Pump
    : a pump is needed to push (i.e., upstream location) or pull (i.e., downstream location) air into and through the animal chamber and respirometry flow-through system.
  • Subsample pump: To pull air through the analyzers, a small, stable, reliable pump is used.

Flow meter
and flow controllers

  • Bubble flow meters: A simple, yet highly accurate way to measure flow rates involves timing movement of bubbles of soap film up glass tubes between marks of known volume.[11] The glass tube is connected at the bottom (for push systems) or at the top (for pull systems) to the air stream. A small rubber pipette bulb attached at the base of the tube acts as both a reservoir and delivery system for the soap bubbles. Operation is simple. First, wet the glass surface along the path bubbles travel (e.g., press the bulb so that copious amounts of soap are pushed up the glass by the air flow) to provide a virtually friction-free surface. Second, pinch the bulb so that one clean bubble is produced. With a stopwatch in hand, record the time required for the bubble to travel between marks on the glass. Note the volume recorded on the upper mark (e.g., 125 = 125 ml), divide the volume by the time required to travel between marks and the result is the flow rate (ml/s). These instruments can be purchased from a variety of sources, but they may also be constructed from appropriate-sized, glass volumetric pipettes.
  • Acrylic flow meters : Under some circumstances of high flow rates we may use simple acrylic flow meters (0–2.5 liters/min) to control the flow rates through the metabolic chambers. The meters are located upstream from the metabolic chambers. The flow meters are simple to use but should be calibrated twice daily for use in the respirometry system: once before recording begins (but after the animal has been sealed inside the chamber!!) and again at the end of the recording (before the animal is removed from the chamber). Calibration must be done with a bubble flow meter because the calibration marks on the acrylic meters are only approximate. For proper calibration of flow rates remember that both barometric pressure and temperature of the air streaming through the flow meter (which we assume to be equal to room temperature) must be recorded.
  • Mass flow meters
    : The equations required for calculating rates of oxygen consumption or carbon dioxide production assume that the flow rates into and out of the chambers are known exactly. We use mass flow meters which have the advantage of yielding flow rates independent of temperature and air pressure. Therefore, these flow rates can be considered to be corrected to standard conditions (Standard Temperature Pressure). We only measure and control flow at one location—downstream from the chamber. Therefore, we must assume that the inflow and outflow rates are identical. However, during construction of the respirometry system, flow rate must be measured at all steps, across all connections, to verify integrity of flow.
  • Needle valves: Mass flow meters may be purchased with mass flow controllers which permit setting flow rates. These are expensive, however. Respirometry research often will attempt to measure more than one animal at a time, which would necessitate one chamber per animal and thus controlled flow through each chamber. An alternative and more cost-effective method to control flow would be via stainless steel or carbon steel needle valves. Needle valves plus mass flow meters provides a cost-effective means to achieve desired flow rates. The valves cost about $20.

Tubing and chambers

  • Tubing and connections : Various kinds of tubing can be used to connect the components of the respirometry system to and from the animal chamber. A variety of kinds of flexible tubing may be used, depending on the characteristics of the system. Acetyl, Bev-A-Line, Kynar, nylon, Tygon tubing and connectors may be used in regions of the system where oxidizing atmospheres are low (e.g., background levels of ozone only);
    Teflon
    tubing would be recommended if there is an expectation for appreciable amounts of ozone to be present because it is inert to ozone. Teflon tubes are more costly and lack flexibility.
  • Metabolic chambers: Chambers may be glass jars with rubber stoppers for lids; syringe barrels for small animals and insects; or constructed from
    Plexiglas. Ideally, chambers should be constructed from inert materials; for example, the acrylic plastics can absorb O2 and may be a poor choice for respirometry with very small insects.[12] Chambers need to be constructed in a manner that yields rapid mixing of gases within the chamber. The simplest metabolic chamber for a small vertebrate might be a glass jar with a stopper. The stoppers are fitted with two ports: short extensions of Teflon tubing are provided for line connections. Teflon tube extensions are pushed through the bulkhead and the line connection is finished by attaching a small hose clip to the base of the Teflon tube extension. Additionally, an extension to the inlet port inside the jar should be provided—this ensures that the animal's expiratory gases are not washed away by the in flow stream. The animal is sealed inside and the rubber stopper is held in place with Velcro
    straps. If an upstream system is used, any metabolic chamber leak will result in loss of animal air and, therefore, an underestimate of the animal's metabolic rate. When you close an animal inside a metabolic chamber, attention must be paid to the seal. To ensure tight seals before closing the lid, firmly work the stopper into the jar and make sure that it is even. Use 1–2 straps (2 are better) and pull tightly. Acrylic (Plexiglas) chambers will be constructed for some uses, but precise engineering will be needed to ensure proper seating; gaskets will help, and judicious use of tight-fitting clamps will minimize leaks.
  • Scrubbing tubes: Water before and after the animal chamber must be removed. One arrangement would use a large acrylic column of
    Drierite (8 mesh (scale), i.e., relatively coarse) upstream (before the push pump, before the animal chamber) to dry incurrent airstream and several tubes with smaller mesh (10–20, i.e., relatively fine) Drierite to remove water after the animal chamber. To prepare a scrubbing tube, make sure there is a small amount of cotton at either end of the tube to prevent dust particles from traveling to the analyzers. Use small amounts of cotton, say around 0.005 g, just enough to keep the dust out of the tubing. Large amounts of cotton will block air flow when/if it gets damp. Pour the Drierite into the tube with a funnel, tap the tube on the bench to pack the grains tightly (to increase surface area – air + water rushes through loose Drierite, requiring frequent changes of scrubbers), and cap off with a small amount of cotton. To remove carbon dioxide] before and after the animal chamber, Ascarite II is used (Ascarite II is a registered trademark of the Arthur H. Thomas Co.). Ascarite II contains NaOH, which is caustic (so don't get any on your skin and keep away from water). A scrubbing tube is prepared by placing a small amount of cotton into the tube end, filling one-third of the way with 10–20 mesh Drierite, adding a small amount of cotton, then an additional third of the tube with the Ascarite II, another layer of cotton, followed by more Drierite and capping the tube off with another small amount of cotton. Tap the tube on the bench as each layer is added to pack the grains. Note: Driereite can be used over and over again (after heating in an oven), although indicating Drierite will lose color with repeated drying; Ascarite II is used once and will be considered a hazardous waste
    .

Analyzers

Finally, a computer data acquisition and control system would be a typical addition to complete the system. Instead of a chart recorder, continuous records of oxygen consumption and or carbon dioxide production are made with the assistance of an analog-to-digital converter coupled to a computer. Software captures, filters, converts, and displays the signal as appropriate to the experimenter's needs. A variety of companies and individuals service the respirometry community (e.g., Sable Systems, Qubit Systems, see also Warthog Systems).

Mitochondrial metabolic rates

Inside the body oxygen is delivered to cells and in the cells to

mitochondria
, where it is consumed in the process generating most of the energy required by the organism. Mitochondrial respirometry measures the consumption of oxygen by the mitochondria without involving an entire living animal and is the main tool to study mitochondrial function.[13] Three different types of samples may be subjected to such respirometric studies: isolated mitochondria (from cell cultures, animals or plants); permeabilized cells (from cell cultures); and permeabilized fibers or tissues (from animals). In the latter two cases the cellular membrane is made permeable by the addition of chemicals leaving selectively the mitochondrial membrane intact. Therefore, chemicals that usually would not be able to cross the cell membrane can directly influence the mitochondria. By the permeabilization of the cellular membrane, the cell stops to exist as a living, defined organism, leaving only the mitochondria as still functional structures. Unlike whole-animal respirometry, mitochondrial respirometry takes place in solution, i.e. the sample is suspended in a medium. Today mitochondrial respirometry is mainly performed with a closed-chamber approach.

Closed-chamber system

The sample suspended in a suitable medium is placed in a hermetically closed metabolic chamber. The mitochondria are brought into defined “states” by the sequential addition of substrates or inhibitors. Since the mitochondria consume oxygen, the oxygen concentration drops. This change of oxygen concentration is recorded by an oxygen sensor in the chamber. From the rate of the oxygen decline (taking into account correction for oxygen diffusion) the respiratory rate of the mitochondria can be computed.[13]

Applications

Basic research

The functioning of mitochondria is studied in the field of bioenergetics.[14] Functional differences between mitochondria from different species are studied by respirometry as an aspect of comparative physiology.[15][16]

Applied research

Mitochondrial respirometry is used to study mitochondrial functionality in

aging.[21]

Equipment

The usual equipment includes a seal-able metabolic chamber, an oxygen sensor, and devices for data recording, stirring, thermostatisation and a way to introduce chemicals into the chamber. As described above for whole-animal respirometry the choice of materials is very important.

PTFE
) may be used. Remaining oxygen diffusion into or out of the chamber materials can be handled by correcting the measured oxygen fluxes for the instrumental oxygen background flux. The entire instrument comprising the mentioned components is often called an oxygraph. The companies providing equipment for whole-animal rspirometry mentioned above are usually not involved in mitochondrial respiromety. The community is serviced at widely varying levels of price and sophistication by companies like Oroboros Instruments, Hansatech, Respirometer Systems & Applications, YSI Life Sciences or Strathkelvin Instruments .

See also

References

  1. ^ White, C. R., and R. S. Seymour. 2005. Allometric scaling of mammalian metabolism. Journal of Experimental Biology 208(9):1611–1619.
  2. ^ Weibel, E. R., and H. Hoppeler. 2005. Exercise-induced maximal metabolic rate scales with muscle aerobic capacity. Journal of Experimental Biology 208(9):1635–1644.
  3. ^ Nagy, K. A. 2005. Field metabolic rate and body size. Journal of Experimental Biology 208(9):1621–1625.
  4. ^ Makarieva, A. M., Gorshkov, V. G., Li, B. L., Chown, S. L., Reich, P. B., & Gavrilov, V. M. (2008). Mean mass-specific metabolic rates are strikingly similar across life's major domains: evidence for life's metabolic optimum. Proceedings of the National Academy of Sciences 105(44):16994-16999.
  5. PMID 2593683
  6. ^ Withers, P. C. 2001. Design, calibration and calculation for flow-through respirometry systems. Australian Journal of Zoology49:445–461.
  7. .
  8. ^ Bartholomew, G. A., D. Vleck, and C. M. Vleck. 1981. Instantaneous measurements of oxygen consumption during pre-flight warm-up and post-flight cooling in Sphingid moths and Saturniid moths. Journal of Experimental Biology90(1):17–32.
  9. ^ Pendar, H., & Socha, J. J. (2015). Estimation of instantaneous gas exchange in flow-through respirometry systems: a modern revision of Bartholomew's Z-transform method. PLOS ONE 10(10): e0139508.
  10. ^ Levy, A. 1964. The accuracy of the bubble meter method for gas flow measurements. Journal of Scientific Instruments 41(7):449–453.
  11. ^ Stevens, E. D. 1992. Use of plastic materials in oxygen-measuring systems. Journal of Applied Physiology 72:801–804
  12. ^
  13. ^ Hildebrandt, T.M. and Grieshaber, M.K., 2008 Three enzymatic activities catalyze the oxidation of sulfide to thiosulfate in mammalian and invertebrate mitochondria. FEBS J. (275): 3352–3361.
  14. ^ Nann A. Fangue N.A., Richards J.G., and Schulte1 P.M. 2009. "Do mitochondrial properties explain intraspecific variation in thermal tolerance?". Journal of Experimental Biology 212:514–522.
  15. ^ Phielix E., Schrauwen-Hinderling V.B., Mensink M., Lenaers E., Meex R., Hoeks J., Kooi M.E., Moonen-Kornips E., Sels J.P., Hesselink M.K., Schrauwen P., 2008 Lower intrinsic ADP-stimulated mitochondrial respiration underlies in vivo mitochondrial dysfunction in muscle of male type 2 diabetic patients. Diabetes 57(11): 2943–9.
  16. ^ Knauf C., Cani P.D., Ait-Belgnaoui A., Benani A., Dray C., Cabou C., Colom A., Uldry M., Rastrelli S., Sabatier E., Godet N., Waget A., Pénicaud L., Valet P., Burcelin R., 2008. Brain glucagon-like peptide 1 signaling controls the onset of high-fat diet-induced insulin resistance and reduces energy expenditure. Endocrinology 149: 4768–4777.
  17. ^ Hoeks J., Briedé J.J., de Vogel J., Schaart G., Nabben M., Moonen-Kornips E., Hesselink M.K., Schrauwen P., 2008. Mitochondrial function, content and ROS production in rat skeletal muscle: effect of high-fat feeding. FEBS Lett. 582: 510–516.
  18. ^ Increase in mitochondrial biogenesis, oxidative stress, and glycolysis in murine lymphomas Enrique Sampera, E., Morgadob, L., Estradab, J.C., Bernadb, A., Hubbarda, A., Susana Cadenas, S. and Melova S., 2009. Increase in mitochondrial biogenesis, oxidative stress, and glycolysis in murine lymphomas. Free Radical Biology and Medicine 46(3): 387–396.
  19. ^ Hutter E., Unterluggauer H., Garedew A., Jansen-Durr P. and Gnaiger E. 2006 High-resolution respirometry – a modern tool in aging research. Exp. Gerontol. 41:103–109.

External links