4D scanning transmission electron microscopy
4D scanning transmission electron microscopy (4D STEM) is a subset of scanning transmission electron microscopy (STEM) which utilizes a pixelated electron detector to capture a convergent beam electron diffraction (CBED) pattern at each scan location. This technique captures a 2 dimensional reciprocal space image associated with each scan point as the beam rasters across a 2 dimensional region in real space, hence the name 4D STEM. Its development was enabled by evolution in STEM detectors and improvements computational power. The technique has applications in visual diffraction imaging, phase orientation and strain mapping, phase contrast analysis, among others.
The name 4D STEM is common in literature, however it is known by other names: 4D STEM
History
The use of diffraction patterns as a function of position dates back to the earliest days of STEM, for instance the early review of John M. Cowley and John C. H. Spence in 1978[2] or the analysis in 1983 by Laurence D. Marks and David J. Smith of the orientation of different crystalline segments in nanoparticles.[3] Later work includes the analysis of diffraction patterns as a function of probe position in 1995, where Peter Nellist, B.C. McCallum and John Rodenburg attempted electron ptychography analysis of crystalline silicon.[4] There is also fluctuation electron microscopy (FEM) technique, proposed in 1996 by Treacy and Gibson, which also included quantitative analysis of the differences in images or diffraction patterns taken at different locations on a given sample.[5]
The field of 4D STEM remained underdeveloped due to the limited capabilities of detectors available at the time. The earliest work used either Grigson coils to scan the diffraction pattern,[6] or an optical camera pickup from a phosphur screen.[7] Later on CCD detectors became available, but while these are commonly used in transmission electron microscopy (TEM) they had limited data acquisition rates, could not distinguish where on the detector an electron strikes with high accuracy, and had low dynamic range which made them undesirable for use in 4D STEM.[8]
In the late 2010s, the development of hybrid pixel array detectors (PAD) with single electron sensitivity, high dynamic range, and fast readout speeds allowed for practical 4D STEM experiments.[9][10]
Operating Principle
While the process of data collection in 4D STEM is identical to that of standard
It is also common for there to be a(n)
The use of a hollow detector with a hole in the middle can allow for transmitted electrons to be passed to an EELS detector while scanning. This allows for the simultaneous collection of chemical spectra information and structure information.
Detectors
In traditional TEM, imaging detectors use phosphorescent scintillators paired with a charge coupled device (CCD) to detect electrons.[11] While these devices have good electron sensitivity, they lack the necessary readout speed and dynamic range necessary for 4D STEM. Additionally, the use of a scintillator can worsen the point spread function (PSF) of the detector due to the electron's interaction with the scintillator resulting in a broadening of the signal. In contrast, traditional annular STEM detectors have the necessary readout speed, but instead of collecting a full CBED pattern the detector integrates the collected intensity over a range of angles into a single data point.[12] The development of pixelated detectors in the 2010s with single electron sensitivity, fast readout speeds, and high dynamic range has enabled 4D STEM as a viable experimental method.[8]
4D STEM detectors are typically built as either a monolithic active pixel sensor (MAPS) or as a hybrid pixel array detector (PAD).[8]
Monolithic active pixel sensor (MAPS)
A MAPS detector consists of a complementary metal–oxide–semiconductor (CMOS) chip paired with a doped epitaxial surface layer which converts high energy electrons into many lower energy electrons that travel down to the detector. MAPS detectors must be radiation hardened as their direct exposure to high energy electrons makes radiation damage a key concern.[13]
Due to its monolithic nature and straightforward design, MAPS detectors can attain high pixel densities on the order of 4000 x 4000. This high pixel density when paired with low electron doses can enable single electron counting for high efficiency imaging. Additionally, MAPS detectors tend to have electron high sensitivities and fast readout speeds, but suffer from limited dynamic range.[14]
Pixel array detector (PAD)
PAD detectors consist of a photodiode bump bonded to an integrated circuit, where each solder bump represents a single pixel on the detector.[9]
These detectors typically have lower pixel densities on the order of 128 x 128 but can achieve much higher dynamic range on the order of 32 bits. These detectors can achieve relatively high readout speeds on the order of 1 ms/pixel but are still lacking compared to their annular detector counterparts in STEM which can achieve readout speeds on the order of 10 μs/pixel.[9][15][10]
Detector noise performance is often measured by its detective quantum efficiency (DQE) defined as:
where is output signal to noise ratio squared and is the input signal to noise ratio squared. Ideally the DQE of a sensor is 1 indicating the sensor generates zero noise. The DQE of MAPS, APS and other direct electron detectors tend to be higher than their CCD camera counterparts.[16][17]
Computational Methods
A major issue in 4D STEM is the large quantity of data collected by the technique. With upwards of 100s of TB of data produced over the course of an hour of scanning, finding pertinent information is challenging and requires advanced computation.[18]
Analysis of such large datasets can be quite complex and computational methods to process this data are being developed. Many code repositories for analysis of 4D STEM are currently in development including: HyperSpy, pyXem, LiberTEM, Pycroscopy, and py4DSTEM.[19][20][21][22][23][8]
These types of AI driven analysis, however, require databases of information to train on which currently do not exist. Additionally, lack of metrics for data quality, limited scalability due to poor cross-platform support across different manufacturers, and lack of standardization in analysis and experimental methods brings up questions of comparability across different datasets as well as reproducibility.[18]
Selected Applications
4D STEM has been utilized in a wide array of applications, the most common uses include virtual diffraction imaging, orientation and strain mapping, and phase contrast analysis which are covered below. The technique has also been applied in: medium range order measurement, Higher order Laue zone (HOLZ) channeling contrast imaging, Position averaged
Virtual Diffraction Imaging
Virtual diffraction imaging is a method developed to generate real space images from diffraction patterns.[8] This technique has been used in characterizing material structures since the 90s[24] but more recently has been applied in 4D STEM applications. This technique is often referred to as scanning electron nano diffraction (SEND) in which 4 pairs of magnetic coils are used to tilt the converged probe and raster the beam across the surface.[25] A "virtual detector," is not a detector at all but rather a method of data processing which integrates a subset of pixels in diffraction patterns at each raster position to create bright-field and dark-field images. The processing of these images is performed by first averaging the diffraction patterns over all scans. Then a region of interest is selected in the diffraction pattern, a virtual aperture is drawn on the reduced dataset, and the image is constructed using computational methods. This virtual aperture can be any size/shape desired and can be created using the 4D dataset gathered from a single scan.[26] This ability to simulate different apertures is enabled by 4D STEM methods and eliminates a typical weaknesses in conventional STEM operation as STEM bright-field and dark-field detectors are placed at fixed angles and cannot be changed during imaging.[27]
With a 4D dataset bright/
4D STEM has been used to map interfaces, select intensity from selected areas of the diffraction plane to form enhanced dark field images,[28] map positions of nanoscale precipitates,[29] create phase maps of beam sensitive battery cathode materials,[30] and measure degree of crystallinity in metal-organic frameworks (MOFs).[31]
Phase Orientation Mapping
Phase orientation mapping is typically done with electron back scattered diffraction in SEM which can give 2D maps of grain orientation in polycrystalline materials.[32] The technique can also be done in TEM using Kikuchi lines, which is more applicable for thicker samples since formation of Kikuchi lines relies on diffuse scattering being present. Alternatively, in TEM one can utilize precession electron diffraction (PED) to record a large number of diffraction patterns and through comparison to known patterns, the relative orientation of grains in can be determined. 4D STEM can also be used to map orientations, in a technique called Bragg spot imaging. The use of traditional TEM techniques typically results in better resolution than the 4D STEM approach but can fail in regions with high strain as the DPs become too distorted.
In Bragg spot imaging, first correlation analysis method is performed to group
Strain Mapping
Utilizing SEND, bright and dark field images can be obtained from diffraction patterns by integration of direct and diffracted beams respectively, as discussed previously. During 4D STEM operation the ADF detector can be used to visualize a particular region of interest through a collection of scattered electrons to large angles to correlate probe location with diffraction during measurements.[25] There is a tradeoff between resolution and strain information; since larger probes can average strain measurements over a large volume, but moving to smaller probe sizes gives higher real space resolution. There are ways to combat this issue such as spacing probes further apart than the resolution limit to increase the field of view.[8]
This strain mapping technique has been applied in many
Phase Contrast Analysis
Differential phase contrast
The differential phase contrast imaging technique (DPC) can be used in
DPC imaging with 4D STEM is up to 2 orders of magnitude slower than DPC with segmented detectors and requires advanced analysis of large four-dimensional datasets.[40]
Ptychography
The overlapping CBED measurements present in a 4D STEM dataset allow for the construction of the complex electron probe and complex sample potential using the ptychography technique. Ptychographic reconstructions with 4D STEM data were shown to provide higher contrast than ADF, BF, ABF, and segmented DPC imaging in STEM. The high signal-to-noise ratio of this technique under 4D STEM makes it attractive for imaging radiation sensitive specimens such as biological specimens[38] The use of a pixelated detector with a hole in the middle to allow the unscattered electron beam to pass to a spectrometer has been shown to allow ptychographic analysis in conjunction with chemical analysis in 4D STEM.[41]
MIDI STEM
This technique MIDI-STEM (matched illumination and detector interferometry-STEM), while being less common, is used with ptychography to create higher contrast phase images. The placement of a phase plate with zones of 0 and π/2 phase shift in the probe forming aperture creates a series of concentric rings in the resulting CBED pattern. The difference in counts between the 0 and π/2 regions allows for direct measurement of local sample phase.[42] The counts in the different regions could be measured via complex standard detector geometries or the use of a pixelated detector in 4D STEM. Pixelated detectors have been shown to utilize this technique with atomic resolution.[43]
(MIDI)-STEM produces image contrast information with less high-pass filtering than DPC or ptychography but is less efficient at high spatial frequencies than those techniques.[8] (MIDI)-STEM used in conjunction with ptychography has been shown to be more efficient in providing contrast information than either technique individually.[44]
See also
- Electron diffraction
- Detectors for transmission electron microscopy
- Energy filtered transmission electron microscopy (EFTEM)
- High-resolution transmission electron microscopy (HRTEM)
- Scanning confocal electron microscopy (SCEM)
- Scanning electron microscope (SEM)
- Scanning Transmission Electron Microscopy (STEM)
- Transmission electron microscopy (TEM)
References
- ^ "4D STEM | Gatan, Inc". www.gatan.com. Retrieved 2022-03-13.
- ISSN 0304-3991.
- S2CID 94822386.
- S2CID 4330017.
- .
- ISSN 0034-6748.
- ^ Cowley, J. M. (1979). "High resolution studies of crystals using STEM". Chemica Scripta. 14: 33–38.
- ^ S2CID 155090551.
- ^ S2CID 5984477.
- ^ PMID 28654827.
- PMID 8266607.
- ^ S2CID 137333314.
- PMID 15890445.
- PMID 25194828.
- S2CID 201288276.
- )
- PMID 25910204.
- ^ PMID 33106651.
- ^ "HyperSpy: multi-dimensional data analysis toolbox — HyperSpy". hyperspy.org. Retrieved 2022-03-13.
- S2CID 214612204.
- ^ "LiberTEM - Open Pixelated STEM platform — LiberTEM 0.10.0.dev0 documentation". libertem.github.io. Retrieved 2022-03-13.
- ^ pyxem/pyxem, pyxem, 2022-03-10, retrieved 2022-03-13
- ^ "Pycroscopy — pycroscopy 0.60.7 documentation". pycroscopy.github.io. Retrieved 2022-03-13.
- ISSN 0022-0744.
- ^ S2CID 207828520.
- ISSN 1477-9986.
- ^ a b "What is 4D STEM?". Blue Scientific. 2020-04-03. Retrieved 2022-03-13.
- S2CID 136699652.
- PMID 19792823.
- PMID 28508061.
- PMID 31189892.
- S2CID 26116915.
- PMID 26028514.
- S2CID 119222107.
- S2CID 139235551.
- PMID 27093600.
- ISSN 1745-2481.
- ^ PMID 25481091.
- S2CID 58006695.
- S2CID 58586726.
- S2CID 53010577.
- ISSN 0304-3991.
- PMID 26923483.
- S2CID 52921918.